Internal Sidestream ETCO2 Module

The C200 OEM ETCO2 module is a complete data collection and analysis system for monitoring respiratory carbon dioxide concentration. The C200 module includes a miniature CO2 sensor, barometric pressure transducer, sampling flow control and a miniature low-power vacuum pump. A microprocessor collects the sensor data and calculates various real-time parameters: instantaneous CO2 concentration, respiration rate, end-tidal CO2, inspired CO2, inspiration and expiration times. The module is highly configurable to meet most OEM implementation requirements. System specifications such as sampling flow rate and floating point data format can be changed by customer. The host sends commands and receives waveform, breath parameters and command responses on two asynchronous serial data lines (TTL or RS232 level)

.

INTERNAL SIDESTREAM ETCO2 MODULE
Model No.: C200


C200 SIDESTREAM ETCO2 MODULE SPECIFICATIONS

Transducer Type

Sidestream CO2 module

Principle of Operation

Nondispersive infrared (NDIR) single beam optics,

dual wavelength, no moving parts

Initialization Time

Capnogram, displayed in less than 10 seconds at

an ambient temperatureof 25ìC,

full specifications within 2 minutes

CO2 Range

0% to 20%, 0 to 152 mmHg

Barometric Pressure supplied by module itself

Rise Time

28 ms (typical) for detector

200ms (typical) for system (Dependent Upon Implementation,

Pneumatics and Water Separation Technique)

CO2 Resolution

0.1 mmHg, 0 to 50 mmHg;

0.2 mmHg, 51 to 100 mmHg;

0.4mmHg, 101 to 152 mmHg

CO2 Accuracy

0 to 40 mmHg ±2 mmHg

41 to 70 mmHg ±5% of reading

71 to 100 mmHg ±8% of reading

101 to 152 mmHg ±10% of reading

Breath Rate

2-150 BPM

Respiration Rate Accuracy

±1 breath

Compensations

Barometric Pressure auto compensated from 400 mmHg to 850mmHg

Operator selectable O2, N2O, HE and Agent compensation.

Calibration

No routine user calibration required. An offset calibration will run

automatically when the ambient temperature is not stable.

Sampling line

Nasal sampling line or extension sampling line

Water Separation

Water trap or gas dryer line

Input Voltage

5.00 Volts (± 5%)

Power Consumption

500mW (typical with pump on)

1600mW (extreme status)

Warm-up Time

10 seconds

Interface

Highly configurable serial digital interfac,(TTL/RS232, CO2 and barometric waveform data,

Breath Detect Algorithm data

Pneumatics

On-board aspiration pump and flow controller

Configurable aspiration flows 50 to 250 ml/min (± 10%)

Temperature and Humidity

Operating: 0 to 50ìC, 10 to 90% RH, non-condensing

Storage: -40 to 70ìC, <90% RH, non-condensing

Dimensions

77mm*52mm*28mm

Weight

< 3.0 ounces (< 76 grams)

Interconnection

Standard 2.54mm, 2*8 pin, female , Pin out:

1 NC; 2 NC; 3 GND; 4 5.0V; 5 EN; 6 NC; 7 TX; 8 NC; 9 RX; 10 NC;

11 NC; 12 5.0V; 13 GND; 14 NC; 15 NC;16 NC.

Note: When EN is NC or connect to high level, module will work,

if EN is connect to GND, module will not work.


HIGH LIGHTS

Best performance price ratio around the world. The price is clearly lower than other

similar module and the performance is better than them.

Better stability. Better design can wipe off the excursion from temperature, pressure

and inconsistent component.

Better selfprotected ability. It allows wrong power supply, humidity in the pipeline.

Better communication interface. TTL or RS232 level can be choose by customer.

Better policy of after service. Manufacturer promises that the module can be replaced

by a new one if it is not damaged by man-made in one year.

Longer life. Using longlife IR source (100 thousand hours life).

The communication protocol of C200 is compatible with Respironics/Philips or CPT